Technical Specifications
Gas Compatibility
Oxygen and oxygen enriched air
Maximum Operating Pressure
600 kPa* *Maximum static pressure delivered by the gas supply source
Flow Rate
0.25* – 25 l/min The Firesafe Cannula Valve can be safely used at lower flow rates but may not meet the internal leak rate of 10 ml/min when activated in the event of a fire.
Resistance to Flow
≤ 0.07 kPa at 2 l/min | ≤ 0.3 kPa at 5 l/min | ≤ 1 kPa at 10 l/min
Intended Life
5 years, expiry date engraved on device
Application
Single patient multiple use
Packaging
100 units per box (5 bags per box, 20 units per bag)
Connections
EN 13544-2, 6 mm nominal tubing connection
Environmental Limits for Transport and Storage
Temperature: -25 °C to 70 °C (-4 °F to 140 °F) Humidity: 0 – 100 % RH
Environmental Limits for Use
Temperature: 0 °C to 50 °C (32 °F to 122 °F) Humidity: 0 – 100 % RH
Applied Standards
EN 13544-2:2002 – Respiratory Therapy Equipment – Part 2: Tubing and connectors.
Regulatory
EC: MDD Class IIa Medical Device, USA: FDA Registered Class 1 Medical Device; 510(k) exempt, GMDN Code and Term: 60391 – Fire safety valve